Restore factory calibration

Restore a sensor's original built-in calibration after it has been manually calibrated.

Note

The images below use a Wireless Temperature Sensor (PS-3201) for demonstration purposes. However, most sensors that can be calibrated with Chemvue will use the same procedure for restoring factory calibration. The only exceptions are sensors with one-click measurement calibrations, such as the Wireless CO2 Sensor (PS-3208) or the Wireless Air Quality Sensor (PS-3226); see the pages for calibrating these sensors for more information.

  1. Connect the sensor to Chemvue.
  2. From any page, select a Displayed Measurement button and select the name of the measurement you wish to restore to its factory calibration.
  3. Select the same Displayed Measurement button again and select Calibrate.

    Calibrate Sensor option

  4. From the dropdown list, select Restore Factory Calibration, then select Next.

    Restore Factory Calibration

  5. Review the factory calibration, then select Restore.

  6. Select OK.