Calibrate an altimeter
Improve the accuracy of an altimeter, such as the one included in the Wireless Barometer (PS-4255) by setting an offset calibration.
- Connect the altitude sensor to Chemvue.
- From any page, select a Displayed Measurement button and select "Altitude" from the list.
-
Select the same Displayed Measurement button again and select Calibrate.
-
Select One Standard (1 point offset) from the dropdown list, then select Next.
-
Enter the sensor's current elevation, as determined via external tools, into the Standard Value box, then select Set Current Value to Standard Value.
-
Review your new calibration to make sure it is appropriate, then select Apply.