Calibrate a Wireless Force-Acceleration Sensor
Improve the accuracy of the Wireless Force-Acceleration Sensor by calibrating its force sensor.
Before you start
Calibrating the force sensor of the Wireless Force-Acceleration Sensor will require the following supplies:
- A 1 kilogram (kg) mass
- The hook attachment provided with the sensor
- A horizontally mounted support rod to hold the sensor
Calibration
- Screw the hook attachment into the Wireless Force-Acceleration Sensor; this will connect the hook to the internal force sensor, allowing force measurements to be collected.
-
Mount the sensor to the support rod as shown in the diagram below. Tighten the thumbscrew to secure the sensor in place.
- Click
in the Tools palette to open the Calibration tool.
- Select Force from the list of measurements to calibrate, then click Next.
-
Select Two Standards (2 point) from the list of calibration types, then click Next.
-
Hang the 1 kg mass from the hook attachment on the sensor.
- The force on the sensor is now -9.8 newtons (N). Enter -9.8 into the Standard Value box, then click Set Current Value to Standard Value.
- Remove the mass from the hook attachment. The force on the sensor is now 0 N.
-
Enter 0 into the Standard Value box, then click Set Current Value to Standard Value.
-
Review the new calibration, then click Finish.